Adjustment and commissioning
The multiCELL associates the two calibrated values to the range of the selected
analogue input signal (slope and offset).
Immerse the probe of the sensor (for insatne a pH sensor) in the first buffer
solution: the multiCELL displays the solution measured value.
Enter the value of the buffer solution (indicated on the bottle).
Wait until the measured value stabilises.
Calibration Mx:Inputs PV calibration
OK
F4
1st point
7.001 pH
7.000 pH
Calibrate using a
2nd point?
NO
2nd point
10.001 pH
10.000 pH
Rinse the probe.
Immerse the clean probe in the second buffer solution: the multiCELL dis-
plays the measured value of the solution.
Enter the value of the second buffer solution (indicated on the bottle).
Wait until the measured value stabilises.
OK
F4
YES
The multiCELL adjusts the
offset.
AI1/AI2
Solution measured value.
Calibration result
Offset: -3,498 pH
Slope:0,876 pH/mA
SAVE
F4
Figure 68 : Example for the calibration of an analogue input with respect to the pH measured by an instrument having a
4-20 mA output
Type 8619