Technical data
Conductivity measurement (with
connected conductivity sensor)
Measurement range
Measurement resolution
Measurement error
0.000 µS/cm...2 S/cm (depends on the conductivity sensor)
10
-9
S/cm
< 0.5% of measured value + sensor error
Resistivity measurement (with connected
conductivity sensor)
Measurement range
Measurement resolution
Measurement error (without sensor)
• 0.500W.cm...100MW.cm (depends on the conductivity sensor)
• 10
-1
W.cm
< 0.5% of measured value
Temperature measurement
Measurement range
Measurement resolution
Measurement error
Temperature sensor type
-40°C ... 200°C
0.1°C
±1°C
Pt100 or Pt1000, with 2 or 3 wires
Electrical connection
• Per detachable 9-point set-screw connector, green
Refer to chap. 6.3.8 for the specifications of the connection
cables
6.3.8. Specifications of the connection cables
Use shielded cables for the electrical connection.
Be sure to use cables in which the wires present the specifications described in the table below.
Wire specifications Dimensions of the max. clamping area
Single core H05(07) V-U 0.2...1.5 mm
2
Flexible wire H05(07) V-K 0.2...1.5 mm
2
With non-insulated end connection 0.2...1.5 mm
2
With insulated end connection 0.2...0.75 mm
2
Table 1: Specifications of the wires making up the connection cables