Installation and wiring
Stage 6:
Fully tighten the screws using an appropriate
screwdriver.
Repeat stages 4 to 6 to fit the other 3 locking
systems.
Figure 4 : Insertion of the 8619 into a housing or cabinet
7.3. Electrical wiring
DANGER
Risk of injury due to electrical discharge.
Shut down and isolate the electrical power source before carrying out work on the system.
Observe all applicable accident protection and safety regulations for electrical equipment.
• Use a high quality electrical power supply (filtered and regulated).
Connect the functional earth on the installation to the ground screw on the device.
Connect the shielding on each wire to an "FE" (functional earth) terminal to guarantee the equipotenti-
ality of the installation.
Use shielded cables that respect the specifications described in "Table 1: Specifications of the wires
making up the connection cables".
7.3.1. Electrical connections
For all versions of the multiCELL, the electrical connection is made by set-screw connectors.
7.3.2. Wiring the M0:MAIN board
The M0:MAIN board is used to connect the multiCELL's electrical power source. It has:
2 digital inputs (marked DI1 and DI2), for connecting a flow sensor for example
Two 4-20 mA analogue outputs (marked AO1 and AO2)
2 digital outputs (marked DO1 and DO2)
The inputs and outputs are galvanically insulated and therefore voltage free.
Type 8619
english