16
Installation and wiring
7.3 Installation of a wall-mounted version
note
Risk of material damage if the cable glands are not tightly screwed on the housing
▶ Before installing the wall-mounted housing on its support, tighten the nuts of the entry item of the cables
glands at a torque of 1.5 Nm.
The flow transmitter in a wall-mounted version has 4 holes in the bottom of the housing.
→ Remove blanking strips covering the screws.
Blanking strips
ENTER
0....9
FLOW
Nut of the entry item
→ Loosen the 4 screws and open the cover to get access to the holes [1].
1 1
230V
LN
230V
T 250 mA
56789 10
LN
230V
CURRENT
SOURCE SINK
BINARY
PE PEPE
PE
DI1
DI2
DI3
DI4
DO4
ISOG
FLOW
SENSOR
L+ L- PE P- P+
NC
Iout
PULSE
DO1
Supply
12..36Vdc
Univ
Batch
(AO1)
SUPPLY
NC
COIL
PULSE
INPUT
NPN/PNP
213PE
+-
SENSOR
SUPPLY
LOAD
+5V
L+
(L+)-12V
COIL/PNP
39K
470
2.2K
SENSOR TYPE
COIL NPN/PNP
DO2
DO3
OFFON
106 mm
Figure 6: Installation of a wall-mounted version
→ Secure the housing to the support respecting the dimensions indicated in “Figure 6”.
→ Wire acc. to instructions in chap. “7.4”.
→ Close the housing and tighten the 4 screws of the cover.
English
Type 8025 UNIVERSAL
Comments to this Manuals